AP Chemistry · Unit 3 · Topic 3.13
3.13 Beer-Lambert Law
TLDR
A spectrophotometer shines one wavelength through a sample and measures how much light gets through. The absorbance A follows A = εbc: ε is the molar absorptivity (how strongly the substance absorbs that wavelength), b is the path length in cm and c is the concentration in M. Twice the concentration or twice the path puts twice as many absorbing particles in the beam, so A doubles. With b and the wavelength fixed, A is proportional to c, so standards make a straight calibration line through the origin with slope εb, and an unknown's concentration is its A divided by that slope. Measure at λmax, the peak of the spectrum, where A changes most.
Guess firstA dye solution lets about 50% of the light through (A = 0.30). You double the dye's concentration. About how much light gets through now?
Pick one now. The page tells you if you were right at the end, after the picture and the sim show you why.
A is correct. Each layer of dye passes the same fraction of the light that reaches it. Twice the dye is like the light crossing the solution twice: half of a half, so about 25% gets through. Absorbance adds instead: 0.30 + 0.30 = 0.60. That is why chemists graph absorbance, not the light through, against concentration.
B is wrong. Trap: that treats the light through as a straight line. The second dose of dye only absorbs a share of the light that is left, so some always gets through.
C is wrong. Trap: ε stays the same, but ε is per particle. Twice the concentration puts twice as many absorbing particles in the beam, so more light is absorbed.
D is wrong. Trap: absorbance is not the percent absorbed. A = 0.30 already means about half the light was absorbed; doubling A takes the light through to about a quarter.
Locked in. Keep reading, the answer is at the end of the page.
The whole topic in one picture
- One wavelength in. The lamp's light goes through a filter that passes only 630 nm, the color this blue dye absorbs most (its λmax). The instrument is first zeroed with a blank, a cuvette of the solvent alone, so 100% means "all the light the dye could take."
- Each slice passes the same fraction. In this solution every 1.00 cm slice holds the same number of dye particles, so each slice absorbs the same share of the light that reaches it: here, half. The beam narrows from 100% to 50% to 25% to 12.5%, a curve, not a straight line.
- Absorbance adds. Each slice adds the same absorbance, 0.30, so 1, 2 and 3 cm give A = 0.30, 0.60 and 0.90. The light left is multiplied (× ½ each time) while A climbs in equal steps: A is proportional to the path length b. That is the b in A = εbc.
- Twice the dye, same path. Put twice the concentration in one 1.00 cm cuvette and the beam meets twice as many particles: 25% gets through and A = 0.60, exactly like two slices of the first solution. A is proportional to c as well.
- The calibration line. With b and the wavelength fixed, A = (εb)c is a straight line through the origin. Standards of known concentration give the line (slope εb = 0.20 per 10−5 M here, so ε = 2.00 × 104 M−1 cm−1). An unknown with A = 0.50 reads off at 2.5 × 10−5 M.
- Absorbance (A)
- How much light a sample absorbs at one wavelength. It has no units and rises in a straight line with the amount of absorbing substance in the beam.
- Molar absorptivity (ε)
- How strongly one substance absorbs one wavelength, in M−1 cm−1. It is fixed for a substance at a given wavelength and is largest at λmax.
- Path length (b)
- The distance the light travels through the solution, in cm. A standard cuvette is 1.00 cm.
- Calibration line
- A graph of absorbance against concentration for standard solutions. With b and λ fixed it is a straight line through the origin with slope εb.
- λmax
- The wavelength where the substance absorbs most, the peak of its absorbance spectrum. Measuring there gives the biggest change in A for each change in c.
- Blank
- A cuvette of the solvent alone, used to zero the instrument so only the solute's absorbance is measured.
What the exam hands you, and what it doesn't
On the exam already
- A = εbc, on the equations sheet
- M = moles of solute ÷ liters of solution, also on the sheet
- The periodic table, for molar masses when the standards are made from a weighed solid
- A calculator for every part of the exam
You bring
- With b and λ fixed, A is proportional to c: double c, double A
- The slope of a calibration line is εb, so c = A ÷ slope
- Pick λmax from the spectrum; a solution absorbs the color it does not show
- Error questions: decide what the mistake does to the measured A first, then to c
- Dilution, M1V1 = M2V2, is not on the sheet: know it to undo a dilution
Try it: Walk the beam through the cell
Set the dye's concentration, the path length and the wavelength. The beam's height shows the light left at every depth, and the graph under it is drawn to the same scale: watch the light fall in a curve while the absorbance climbs in a straight line. Switch the graph to see the calibration line.
Work it: Build the line, read the unknown, judge a mistake
Choose the wavelength, find the slope and ε from the standards, read the unknown's concentration, then decide how a lab mistake moves the result. A wrong answer gets a reason, and a second miss shows the line.
How it shows up on the exam
- Calculation from A = εbc or from a calibration line: find c from A, find ε from a slope, or predict A for a new path length. Keep b in cm, because ε is per cm.
- Reading a spectrum or a set of standards: pick the best wavelength (λmax), read an unknown off the line, or undo a dilution to get the original concentration.
- Lab error, usually on free response: "Would the calculated concentration be greater than, less than or equal to the actual value? Justify your answer." The point needs a direction and the reason through A: for example, a wet cuvette dilutes the sample, so fewer particles are in the beam, the measured absorbance is lower and A = εbc gives a lower c.
Find the wrong step
A sports drink is diluted from 2.00 mL to 10.0 mL with water and measured in a 0.500 cm cuvette: A = 0.450, ε = 1.50 × 10⁴ M⁻¹ cm⁻¹. A student finds the dye concentration in the drink. Click the line with the mistake.
The mistake: The drink is more concentrated than the diluted sample in the cuvette, not less. Undoing a fivefold dilution means multiplying by 5 (M₁V₁ = M₂V₂: M₁ = 6.00 × 10⁻⁵ M × 10.0 mL ÷ 2.00 mL).
The drink was diluted 5 times, so c(drink) = 6.00 × 10⁻⁵ M × 5 = 3.00 × 10⁻⁴ M.
Practice: 3 questions
Original AP-style questions. Every wrong choice is a real trap, and the explanation names it.
Cuvette X holds a 2.0 × 10−5 M dye solution with a 1.00 cm path. Cuvette Y holds a 1.0 × 10−5 M solution of the same dye with a 2.00 cm path. Both are measured at the same wavelength. Which statement is correct?
A is wrong. Trap: A depends on the path length too. Y's beam crosses twice as much solution, which makes up for the lower concentration.
B is wrong. Trap: b and c count exactly the same in A = εbc. Each one doubled doubles A; here one doubled and the other halved.
C is correct. A = εbc. For X, b × c = 1.00 × 2.0 × 10−5; for Y, 2.00 × 1.0 × 10−5. The products are equal and ε is the same, so the absorbances are equal: half the particles per cm, over twice the distance.
D is wrong. Trap: the instrument is zeroed with a solvent blank, and the solvent does not absorb at this wavelength. Only the dye particles in the beam count.
Standards of a purple dye measured at 560 nm in a 1.00 cm cuvette give: 1.00 × 10−4 M, A = 0.150; 2.00 × 10−4 M, A = 0.300; 3.00 × 10−4 M, A = 0.450; 4.00 × 10−4 M, A = 0.600. An unknown in the same cuvette gives A = 0.390. What is its concentration?
A is wrong. Trap: that multiplies A by the slope. A = (εb)c, so c is A divided by the slope. No solution is hundreds of molar.
B is wrong. Trap: that reads the absorbance as if it were the concentration in units of 10−4 M. Here 0.150 of A matches 1.00 × 10−4 M, not 0.100.
C is wrong. Trap: that is 1 ÷ slope, which leaves out the unknown's absorbance. It is also beyond the top standard, though A = 0.390 is inside the range.
D is correct. The line goes through the origin with slope 0.600 ÷ 4.00 × 10−4 M = 1.50 × 103 M−1. c = A ÷ slope = 0.390 ÷ 1.50 × 103 M−1 = 2.60 × 10−4 M, between the second and third standards.
A student makes a calibration line with standards measured at 630 nm, the dye's λmax. By mistake, the unknown is measured at 590 nm, where the dye's ε is smaller. How does this affect the calculated concentration of the unknown?
A is wrong. Trap: less light absorbed means a lower absorbance, and a lower A reads off the line as a lower concentration, not a higher one.
B is correct. At 590 nm each dye particle absorbs less (smaller ε), so the unknown's measured A is smaller than the A it would give at 630 nm. Read against the 630 nm line, the smaller A gives a smaller c than the true value.
C is wrong. Trap: the real concentration does not change, but the calculated one does. The unknown's A depends on ε, and it was compared with a line made at a different ε.
D is wrong. Trap: it holds at each wavelength with that wavelength's ε. A line made at 630 nm only works for readings made at 630 nm.
How it connects
Ideas from this topic show up again in other topics. These are the links the exam uses most.
Back to your guessA dye solution lets about 50% of the light through (A = 0.30). You double the dye's concentration. About how much light gets through now?
Answer: A. 25%: the absorbance doubles to 0.60. Each layer of dye passes the same fraction of the light that reaches it. Twice the dye is like the light crossing the solution twice: half of a half, so about 25% gets through. Absorbance adds instead: 0.30 + 0.30 = 0.60. That is why chemists graph absorbance, not the light through, against concentration.
Quick questions
What does each letter in A = εbc stand for?
A is the absorbance (no units), ε is the molar absorptivity in M⁻¹ cm⁻¹ (how strongly the substance absorbs that wavelength), b is the path length in cm, and c is the concentration in mol/L. The equation is on the AP equations sheet.
Why do you measure absorbance at λmax?
At the wavelength of maximum absorbance, ε is largest, so a small change in concentration gives the largest change in absorbance. The measurement is most sensitive there, and small errors in the reading matter least.
Is absorbance the same as the percent of light absorbed?
No. Absorbance rises in a straight line with concentration, but the light that gets through falls in a curve: A = 0.30 lets about 50% through, A = 0.60 about 25%, A = 1.0 about 10%. That is why calibration lines graph absorbance, not percent transmitted.
Go deeper
Light and Spectroscopy: the full Unit 3 lesson
About 13 minutes of interactive slides on topics 3.11 to 3.13, with step-by-step math, particle drawings, a diagnostic that finds your weak spots, a 50-question bank and AI-scored free response.